<Error>
<ErrorNumber>48006</ErrorNumber>
<ErrorFamily>Media Handler</ErrorFamily>
<ErrorMessage>
MDH: Shuttle failed to move media to the loadtable, check &lt;Media location&gt;
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The shuttle attempted to move media to the loadtable however media was still detected
on the shuttle.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Check the device to ensure media is not caught or blocked
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Faulty hardware.
</Cause>
<Cause>
Faulty sensors (ST_SensorPinchRollersHomePlateDetect, ST_SensorPinchRollersAwayPlateDetect)
</Cause>
<Parameter>
Media location -
Location in the device where the media was found
</Parameter>
<LastModified>
03-Dec-2007
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:53 2008 PST
</Updated>
<Links>
</Links>
</Error>